SOP for Operation of Sampling Booth

0
139

Standard operating procedure for Sampling Booth to avoid airborn contamination in material being sampled.

1.0 OBJECTIVE
To avoid airborne contamination in the material during sampling.

2.0 SCOPE
This procedure is applicable during sampling of raw material in sampling booth installed in the raw material store.

3.0 RESPONSIBILITY
3.1 Doing: Technical Assistant
3.2 Checking: Executive/Manager

4.0 ACCOUNTABILITY
Head of the Department

5.0 PROCEDURE

5.1 Introduction
5.1.1 It is important and most essential to avoid any air born contamination in the material during sampling.
5.1.2 To sample the raw material under “Clean environment” the sampling booths are designed in which fresh air is initially sucked through core filter then 5 micron filter, before passing through HEPA filter to reduce the load on HEPA filter.
5.1.3 Core filter and 5 micron filter are cleanable and require cleaning it periodically as and when required.
5.1.4 The recommended manometer reading is between 7 to 15 mm of water.
5.1.5 If reading goes more than 15, it indicates chocking of HEPA filter and needs replacement of HEPA filter.
5.1.6 Whenever replacement of HEPA filter takes place, the system is to be validated w.r.t. PAO test.
5.1.7 If reading goes down, it indicates chocking of prefilter which requires cleaning. Inform engineering department for cleaning of prefilter.

5.2 Operation of sampling booths
5.2.1 Switch ‘ON’ the mains.
5.2.2 Switch ‘ON’ the light.
5.2.3 Start blower at slow speed one hour before the sampling.
5.3 Note the reading of differential pressure daily.
5.4 Always open material for sampling when the system is ‘ON’
5.5 Sample the material as per raw material sampling SOP.
5.6 Don’t take sampling device outside of the booth, but collect in a closed container in booth itself, to avoid any spillage outside the booth.
5.7 After sampling every raw material, note the manometer reading.
5.8 Use clean sampling device for different material.
5.9 Close the polybag/ containers of sampled material under laminar flow unit, before taking it out.
5.10 Put ‘OFF’ the system after use.

6.0 ABBREVIATIONS
6.1 HEPA = High Efficiency Particulate Air
6.2 PSIG = Pound per Square Inch Gauze
6.3 DOP = Di Octyl Pthalate